Labo C-Top
REF 14876
Benchtop incubator for up to date,
individual cell cultivation
Advantages of Labo C-Top:
- Approved medical device
- Two independent and separate incubation chambers;
Parameters (gas flow, temperature) can be adjusted individually - Optimized gassing with premixed gas
- Optimized temperature distribution by double heated chamber from top and bottom;
Very fast heating up and recovery times - Reliable humidity system;
Current value of relative humidity is displayed (% rH) - Convenient glass control panel
- Space saving
Incubation chambers | Aluminum milled and eloxated In the lower part:
Independent, for use with (per chamber):
|
Temperature | Overall heating from top and bottom Very homogeneous temperature distribution |
Humidity | Humidification by means of humidity block Current value of relative humidity is displayed (% rH) |
Gassing | Connection of premixed gas (CO2, O2 < 21 %, N2) Separate measuring and regulation of gas flow for each chamber Automatic gas purge when closing the chamber |
Recovery times | Short recovery times for all parameters through microprocessor controlled regulation |
Use of glass panel | Homogeneous glass cover Sensoric buttons integrated in glass Image of placements for marking inserted dishes at the glass cover Easy operation and menu navigation Lit LCD at working place height |
Trouble check system | Optical and acoustic alarm in case of set point deviation or defective sensor Independent over-temperature protection Remote alarm |
Options | PC software LaboDat+ for documentation of incubator’s parameters Second gas inlet for use of different gas mixtures for each chamber Lockable lids |
Interfaces | Remote alarm (potential free) RS485-interface for data logging with optional available PC software LaboDat+ |
Exterior | Dimensions 403 x 225 x 155 mm (w x d x h) Weight 8.8 kg |
Interior | Two chambers with 0.5 liters each Mixed load possible |
Temperature | Overall heating from top and bottom of each chamber Range 30 °C - 42 °C from 7 °C above ambient temperature Stability / uniformity ± 0.1 °C / ± 0.3 °C |
Gassing | Premixed gas Gas flow adjustable between 20 ml/min and 30 ml/min Flow rate after opening the lid (automatic gas purge): 250 ml/min for 3 minutes |
Power Supply | 110 – 230 V AC, 50/60 Hz, 100 VA Primary pressure for gas 0.8 bar Ambient temperature 18 °C - 30 °C |
Classification | Safety class II Class IIa for every usage according to EC-Directive 93/42/EEC EN 61010 conform |
REF 15070 Labo C-Top with lockable lids Limited access to the cultures
| |
REF 15069 Labo C-Top with second gas inlet
Every chamber to be gassed with different mixtures at the same time |