CO₂ Incubator Labo C201
CO₂ Incubator Labo C201
REF 15245
Incubator with 200 liter chamber capacity
Advantages of Labo C201:
- 7“ colour touch screen
- UV decontamination (option)
- Particle filter (P3) within airflow
- Graphical display of recorded parameters of the past 2 h or 24 h (T, CO2, rH, O2)
- Cable bushing (Access port)
- Access control via numerical code (option)
- RS485 or ethernet interface
- Active sterile humidity:
- CO2 sensor:
- Direct heating:
- Optional:
Interior | Electro-polished stainless steel with coved corners Shelves and ductwork easily removable for cleaning purpose Access port Ø 30 mm |
Temperature | Intelligent temperature control system avoids condensation in inner chamber |
Humidity supply | Active sterile humidity via external water supply container |
CO2 control | Dual beam infrared sensor Measurement independent from temperature and humidity |
Fast recovery | For all adjustable parameters through optimized microprocessor control |
Easy handling | Menu guided 7“ colour touch display Graphical display of recorded parameters of the past 2 h or 24 h |
Diagnostic system | Visual and acoustic alarm if set point deviation or sensor defect occurs Door monitoring Remote alarm |
Options | O2 control: connection for N2 Subdivided inner glass door with 6 small doors Access control via numerical code UV decontamination routine PC software LaboDat+ |
Interfaces | Alarm contact (potential free) RS485 or Ethernet interface |
Exterior | Dimensions 720 x 700 x 870 mm (w x d x h) |
Interior | Chamber volume 200 l Electro-polished stainless steel Shelves height adjustable |
Temperature | Panel heated interior chamber and door Range 27-42 °C starting 5 °C above ambient temperature Stability / uniformity ±0.1 °C / ±0.3 °C Over-temperature alarm 32-47 °C |
CO2 control | Dual beam infrared sensor Range 0–10 % CO2 Stability ±0.1 % CO2 |
O2 control < 21 % (optional) | Measurement with galvanic sensor Connection of N2 Range 1-21 % O2 |
Humidity | Active sterile humidity generated by vaporization module at 120 °C Measuring range 0–98 % rH / adjustment range 60 % – 95 % rH |
Access port | Ø 30 mm |
Supply | 230 V AC, 50/60 Hz or 115 V AC, 50/60 Hz, 250 VA Inlet pressure for gas 0.8 bar Ambient temperature 18-30 °C |
Classification | Safety class I Conform to EN 61010 |
Subdivided inner glass door - for even faster recovery 6 small inner doors Even faster recovery of all parameters Even lower gas consumption Easy cleaning of inner doors Scope of supply:
| |
UV decontamination Individual start by user Duration: 180 minutes Short-wave UV radiation (UV-C radiation, <280 nm) UV mode is terminated immediately when the door is openened
| |
Access control - higher security for your cultures
Mechanical unlocking in case of power failure (emergency release)
| |
Stacking kit - for safe positioning
Safe, easy and space-saving positioning of the devices in the laboratory
REF 10681 High frame stand, 115 cm height, white coated steel
| |
Roller base
For mobile positioning of the devices in the laboratory, on lockable wheels, preventing roller base movement
| |
Our Labo C201 incubators are available with O2-control (O2 < 21 %) | |
Each Labotect CO2 incubator passes an individual cell-growth test prior to delivery! | |
Further accessories of the incubator Labo C201 |
back